HORIBA OCMA-550 Oil Content Analyzer

  • Sale
  • $6,616.60
  • Regular price $7,399.95
Shipping calculated at checkout.



OCMA-550 is used to measure residual oil on components and concentrations of oil adhered on solids such as soil. Measurement can be easily made only by injecting the extracted water sample into the attached cell and setting it to the equipment.

This model features a simple design which allows opening/closing of the door to setting of cell with just one hand. This is best for measurement of extracted samples such as evaluation of residual oil on components and measurement of oil contained in food.

 

FEATURES AND BENEFITS

  • Any oil with low boiling point can be measured

    • With no need for a solvent evaporation process, the OCMA-500 series offers easy operation, and can detect even oils with low boiling points
  • Cell is easily detactable with just one hand

    • Simple design which allows opening/closing of door and detachment of cell with just one hand. Measurement operation becomes smoother.
  • Timer function
    • A timer function to display a measurement value in a certain amount of time is equipped. This saves work and time required for measurement.
  • Color graphic LCD
    • The OCMA-500 series comes equipped with a 3.5-in color graphic LCD. Menus and measurement data can be clearly see
  • USB flash memory drive port
    • Storing data on USB flash memories enables easy data management on PCs.
  • Unit conversion
    • Inputting measurement conditions can change the units displayed (mg/L, mg/kg, mg/g, and mg/PC) as the user needs.

         

        APPLICATIONS

        • Residual oil components
        • For piping of air conditioners and refrigerator
        • Semiconductor field
        • Useful for soil (environmental pollution), food (health hazard) and gas (quality deterioration)

         

        SPECIFICATIONS

        Measurement Principle Solvent extraction-non-dispersive infrared absorption analysis method
        Measurement Range 0mg/L ~ 200mg/L
        Resolution 0 to 99.9 mg/L; 0.1 mg/L
        100 to 200 mg/L; 1 mg/L
        Repeatability 0 to 9.9 mg/L; ± 0.4 mg/L ± 1 digit
        10.0 to 99.9 mg/L; ± 2 mg/L ± 1 digit
        100 to 200 mg/L; ± 4 mg/L ± 1 digit
        *For standard liquids 
        Display Method 3.5 inches, 320×240 dots
        Backlight Colour graphic LCD
        Calibration Method Select each optionally zero and span calibration
        Extraction Solvent S-316 solvent
        ※Do not use any other solvent than S-316
        Amount of Extraction Solvent Required Approx. 6.5ml (Amount of extration solvent required)
        Extraction Method Using the extraction solvent and extracted manually outside the product
        Ambient Operationg Temperature 0 to 40 ℃ (non condensation)
        Power Supply AC 100V to 240V ±10%, 50/60 Hz
        Power Consumption AC 100V to 240V : Approx.60VA
        External Dimensions 195(H)×253(W)×293(D) mm
        Mass Approx. 5 kg
        External Output Output to an USB memory stick
        Functions 300-item data memory(Measurement history)
        Stabilized measurement value display
        Self error determination
        Clock function
        Unit conversion function