HORIBA OCMA-500 Oil Content Analyzer
by Horiba
Save 7%
Original price
$10,900.00
Current price
$10,100.00
The OCMA-500 Oil Content Analyzer has been redesigned for even better operability, maintaining its user-friendly features. After injecting the sample, all you have to do is press a button to get the monitoring operation done quickly, from oil extraction to sample measurement and draining. This machine is easy for anyone to use.The OCMA-500 cuts solvent consumption by 20% compared with our previous products, reducing environmental impact. It also reduces the running costs.
FEATURES AND BENEFITS
- Fully automated one-touch operation, from oil extraction to sample measurement and draining
- Reduced environmental impact and running costs (20% cut in solvent consumption)
- With no need for a solvent evaporation process, the OCMA-500 offers easy operation, and can detect even oils with low boiling points
- The extraction tank is equipped with LEDs. Illuminating the tank makes it easy to check the phase separation between sample and solvent
- The OCMA-500 comes equipped with a 3.5-in color graphic LCD. Menus and measurement data can be clearly seen
- Storing data on USB flash memories enables easy data management on PCs
- Inputting measurement conditions can change the units displayed (mg/L, mg/kg, mg/g, and mg/PC) as the user needs.
APPLICATIONS
- Factory drainage
- Ballast water, Bilge
- Vehicle mentenance site and gas station
- Water quality survey based on environmental standard
- For monitoring final discharge from petroleum refinery plant
- For oil dispersion reserch at time of an accident
- For monitoring final discharge water
SPECIFICATIONS
Measurement Principle | Solvent extraction, NDIR analysis (Infrared spectroscopy) |
Measurement Range | 0mg/l ~ 200mg/l dynamic ranging |
Resolution | 0 to 99.9 mg/L; 0.1 mg/L 100 to 200 mg/L; 1 mg/L |
Repeatability | 0 to 9.9 mg/L; ± 0.2 mg/L ± 1 digit 10.0 to 99.9 mg/L; ± 2 mg/L ± 1 digit 100 to 200 mg/L; ± 4 mg/L ± 1 digit *For standard liquids |
Display Method | 3.5 inches, 320×240 dots Backlight Colour graphic LCD |
Calibration Method | Zero, Span Calibration |
Extraction Solvent | S-316 solvent |
Amount of Test Sample Required | Sample water : Solvent = 2 : 1 |
Amount of Extraction Solvent Required | 8 mL |
Extraction Method | Built-in extractor |
Ambient Operationg Temperature | 0 to 40 ℃ (non condensation) |
Power Supply | AC 100V to 240V ±10%, 50/60 Hz |
Power Consumption | AC 100V:Approx.60VA AC 240V: Approx.90VA |
External Dimensions | 200(W)×313(D)×342(H) mm |
Mass | Approx. 7 kg |
External Output | Output to an USB memory stick |
Functions |
300-item data memory |