HORIBA OCMA-500 Oil Content Analyzer

  • Sale
  • Regular price $10,349.95
Shipping calculated at checkout.


The OCMA-500 Oil Content Analyzer has been redesigned for even better operability, maintaining its user-friendly features. After injecting the sample, all you have to do is press a button to get the monitoring operation done quickly, from oil extraction to sample measurement and draining. This machine is easy for anyone to use.The OCMA-500 cuts solvent consumption by 20% compared with our previous products, reducing environmental impact. It also reduces the running costs.

 

FEATURES AND BENEFITS

  • Fully automated one-touch operation, from oil extraction to sample measurement and draining
  • Reduced environmental impact and running costs (20% cut in solvent consumption)
  • With no need for a solvent evaporation process, the OCMA-500 offers easy operation, and can detect even oils with low boiling points
  • The extraction tank is equipped with LEDs. Illuminating the tank makes it easy to check the phase separation between sample and solvent
  • The OCMA-500 comes equipped with a 3.5-in color graphic LCD. Menus and measurement data can be clearly seen
  • Storing data on USB flash memories enables easy data management on PCs
  • Inputting measurement conditions can change the units displayed (mg/L, mg/kg, mg/g, and mg/PC) as the user needs.

 

APPLICATIONS

  • Factory drainage
  • Ballast water, Bilge
  • Vehicle mentenance site and gas station
  • Water quality survey based on environmental standard
  • For monitoring final discharge from petroleum refinery plant
  • For oil dispersion reserch at time of an accident
  • For monitoring final discharge water

 

SPECIFICATIONS

Measurement Principle Solvent extraction, NDIR
analysis (Infrared spectroscopy)
Measurement Range 0mg/l ~ 200mg/l dynamic ranging
Resolution 0 to 99.9 mg/L; 0.1 mg/L
100 to 200 mg/L; 1 mg/L
Repeatability 0 to 9.9 mg/L; ± 0.2 mg/L ± 1 digit
10.0 to 99.9 mg/L; ± 2 mg/L ± 1 digit
100 to 200 mg/L; ± 4 mg/L ± 1 digit
*For standard liquids 
Display Method 3.5 inches, 320×240 dots
Backlight Colour graphic LCD
Calibration Method Zero, Span Calibration
Extraction Solvent S-316 solvent
Amount of Test Sample Required Sample water : Solvent = 2 : 1 
Amount of Extraction Solvent Required 8 mL
Extraction Method Built-in extractor
Ambient Operationg Temperature 0 to 40 ℃ (non condensation)
Power Supply AC 100V to 240V ±10%, 50/60 Hz
Power Consumption AC 100V:Approx.60VA
AC 240V: Approx.90VA
External Dimensions 200(W)×313(D)×342(H) mm
Mass Approx. 7 kg
External Output Output to an USB memory stick
Functions

300-item data memory
Stabilized measurement value display
Self error determination
Clock